Following the introduction of the Laryngeal Mask Airway (LMA) Classic™ (LMA™ North America, Inc., San Diego, CA, USA) in the 1980 s, the use of laryngeal mask airway devices in elective and emergent airway management has become widespread in clinical practice. The role of the LMA Classic during general anesthesia, both as a ventilation device and as a conduit for intubation, has been amply described in literature.1-3 Since then, newer supraglottic airway models have been introduced. The LMA Proseal™, (LMA™ North America, Inc., San Diego, CA, USA) developed in 2000, consists of a double lumen which separates the respiratory and alimentary tracts, improving protection from aspiration.4-6 Compared with the LMA Classic, the modified inflatable cuff of the LMA Proseal has higher oropharyngeal leak pressure around the laryngeal outlet.7 It has also been used as an acceptable alternative to the endotracheal tube in situations with potential gastric aspiration, such as in Cesarean delivery or laparoscopic surgery.8,9 The LMA Proseal is thus considered the standard “state of the art” supraglottic airway, and it should be used as the standard of comparison for newly developed supraglottic airways.

The LMA Supreme™, (LMA™ North America, Inc., San Diego, CA, USA) developed in 2007, is a modified single-use version of the LMA Proseal. Its preformed curved shaft consists of a double lumen, i.e., a central lumen for access to the digestive tract encased within a flattened oval-shaped airway lumen for access to the respiratory tract.10-13 Other features of the LMA Supreme include a built-in bite block to prevent airway obstruction due to patient bite and moulded fins at the laryngeal outlet to prevent epiglottic obstruction (Fig. 1).

Fig. 1
figure 1

Frontal (left) and side (right) views of the LMA Supreme. ET = epiglottic fins; FT = fixation tab

The LMA Supreme is inserted with the cuff fully deflated, and it is recommended that lubrication be applied to the posterior surface of the airway and cuff to facilitate passage into the airway. With the patient positioned in a semi-sniffing position, the device is advanced with the distal tip flat along the hard palate. The device is then rotated inwards using a single hand, similar to the LMA Fastrach™, until resistance is felt. If correctly positioned, the tip of the cuff should rest at the esophageal inlet. The cuff is then inflated to the standard intracuff pressure of 60 mmHg.11,14 The LMA Supreme is usually fixated by taping across the fixation tab, with the tab situated 1-2.5 cm from the patient’s upper lip. If the fixation tab is < 1 cm or > 2.5 cm from the upper lip, this suggests that the size of airway device is either too small or too big, respectively. Since its release, the LMA Supreme has been used both as a supraglottic airway for ventilation during general anesthesia and as a conduit for tracheal intubation. This article is a review of the efficacy and safety of the LMA Supreme as a stand-alone device and as a conduit for tracheal intubation.

We used the MEDLINE (1948-July 2011) and EMBASE (1980-July 2011) databases to find potential articles, and the keywords “LMA Supreme” and “sLMA” were used in the search. Nineteen articles were found, 12 of which reported randomized controlled trials (RCTs) comparing the LMA Supreme with other airway devices; seven articles were prospective single-device studies and four were case reports. We assessed a number of outcomes of interest, including insertion success, insertion time, oropharyngeal leak pressure, and complications. The characteristics and outcomes of the studies are listed in Tables 1 and 2, respectively.

Table 1 Randomized controlled trials and prospective single-device studies evaluating the LMA Supreme™
Table 2 Randomized controlled trials and prospective single-device study outcomes

Use of LMA Supreme in normal airway/elective surgery

Comparison of the LMA Supreme and LMA Proseal

We located 12 published RCTs examining use of the LMA Supreme as a stand-alone supraglottic airway (Table 1). Six of the studies compared the LMA Supreme with the LMA Proseal.10,13-17 The remaining studies compared the LMA Supreme with the LMA Classic,23 i-gel,18-20 facemask,21 or endotracheal tube.22

In the six RCTs comparing the LMA Supreme with the LMA Proseal, 418 patients were identified and 295 (70%) of those were female. Three of the six trials were parallel randomized controlled trials, while the other three were crossover RCTs. All patients were undergoing elective surgery, and those with an anticipated difficult airway were excluded from the study. A summary of the study outcomes are listed in Table 2. Success rates were similar between the two airway devices in all studies. Overall insertion success rates were 100% for both the LMA Supreme and LMA Proseal. First attempt success rates were similar and ranged from 90-98% for the LMA Supreme and 83-97% for the LMA Proseal.

Insertion end times were defined differently in each study. End points included placement of the LMA,13 obtaining effective airway/ventilation,14,17 connection to the anesthesia circuit,10 and obtaining a satisfactory capnography tracing.15 Despite subtle differences in end points, insertion times were similar between the LMA Supreme and LMA Proseal in all studies.

Oropharyngeal leak pressure associated with the LMA Supreme and LMA Proseal was assessed. In three studies, oropharyngeal leak pressure with the LMA Proseal was significantly higher than that with the LMA Supreme,13,15,17 but it did not differ significantly in the other three studies.10,14,16 In a study by Hosten et al., leak pressure was comparable between groups at one minute and at 60 min after insertion of the airway device; however, the pressure within both groups at 60 min was significantly lower compared with the pressure at one minute.14

Five of the studies recorded complications, including blood staining on the airway device and sore throat (Table 2).13-17 Complications were infrequent and did not differ appreciably between devices. Overall, the incidence of blood staining was 18/207 (9%) with the LMA Proseal compared with 13/208 (6%) with the LMA Supreme. Lee et al. also reported just three occurrences of sore throat for both devices and one case of hoarseness with the LMA Proseal.17

Comparison of LMA Supreme with other airway devices

In six randomized controlled studies, the LMA Supreme was compared with other airway devices, including the face mask, endotracheal tube, LMA Classic, and i-gel.

Comparison of the LMA Supreme and i-gel

Three RCTs compared the LMA Supreme with i-gel.18-20 Two hundred fifty patients were identified and 191 (76%) of those were female. All patients were undergoing elective surgery. Two studies compared the devices in situations of a predicted normal airway,19,20 but in the third study, the devices were assessed in the context of a difficult airway through simulation using an extrication collar.18 The first attempt and overall success rates were high but not significantly different, ranging from 93-98% and 85-96% on first attempt and 94-100% and 93-100% overall with the LMA Supreme and i-gel, respectively.

Airway insertion time was defined differently in each study. Ranges included time from removal of the facemask to successful ventilation,18 from picking up the device to obtaining the first end-tidal CO2 tracing,19 and time from insertion of the device to first end-tidal CO2 tracing.20 In one study, insertion times were faster with the LMA Supreme than with the i-gel [34 (12) sec vs 42 (23) sec; P = 0.024],18 but they were not significant in the other two studies.19,20

In one study, the oropharyngeal leak pressure was found to be higher with the LMA Supreme than with the i-gel,19 but it was not significant in the other studies.18,20 Complications were infrequent with both devices. Due to the crossover nature of Theiler et al.’s study, postoperative discomfort could not be determined.18

Comparison of the LMA Supreme with the facemask

The LMA Supreme was compared with the facemask in 31 morbidly obese patients with difficult facemask ventilation predictors.21 After brief training, ten medical students, novel to airway management, performed the facemask and LMA Supreme airway management techniques. Ventilation using the facemask was successful in 27/31 (87%) cases while ventilation using the LMA Supreme was successful in 31/31 (100%) cases. The insertion time, from picking up each device to the first CO2 tracing, was significantly higher with the ventilation facemask than with the LMA Supreme [34 (14) sec vs 21 (9) sec, respectively; P < 0.05].

Comparison of the LMA Supreme and endotracheal tube

The LMA Supreme was also compared with the endotracheal tube.22 One hundred thirty-eight female patients undergoing elective surgery were randomized into either device arm, and there were no insertion failures with either device. Time to successful airway placement, from injection of atracurium to definitive ventilation, was 3.8 (0.7) min for the endotracheal tube and 2.2 (0.6) min for the LMA Supreme (non-significant). Oropharyngeal leak pressure was > 30 cm H2O in 95% of cases with the LMA Supreme and in 100% of cases with the endotracheal tube. Postoperative discomfort, such as hoarseness of voice, dysphagia, and sore throat were significantly higher in the endotracheal tube group (P < 0.05).

Comparison of the LMA Supreme and LMA Classic

The LMA Supreme was compared with the LMA Classic in 70 patients undergoing surgery.23 Insertion success on first attempt was 27/35 (77%) using the LMA Classic and 31/35 (88%) using the LMA Supreme, and overall insertion success was 31/35 (89%) and 34/35 (97%), respectively (non-significant). Insertion time was faster with the LMA Supreme than with the LMA Classic [8 (1.7) sec vs 11.2 (2.5) sec, respectively; P < 0.001]. Oropharyngeal leak pressure was also higher with the LMA Supreme than with the LMA Classic [34.6 (3.1) cm H2O vs 26.1 (2.1) cm H2O, respectively; P < 0.001]. An adverse effects assessment for nausea and vomiting, agitation, and sore throat was conducted one hour post-operation, and no statistical significance was found between the two groups.

Thus, the LMA Supreme was shown to be comparable with the LMA Proseal with regard to airway success rate, device insertion time, and a favourable complication profile. However, in three of the six studies, oropharyngeal leak pressure was significantly higher in the LMA Proseal than in the LMA Supreme. When compared with the i-gel, the LMA Supreme was comparable with respect to success rates and complications, but some studies indicated a shorter insertion time and higher oropharyngeal leak pressure compared with the LMA Supreme. There are insufficient data to formulate conclusions when comparing the LMA Supreme with the facemask, endotracheal tube, and LMA Classic.

LMA Supreme: prospective single-device studies

The LMA Supreme has been investigated in seven prospective controlled single-device studies involving 507 patients undergoing general anesthesia 11, 12, 2428 (Table 2). Together, these studies showed overall insertion success in 616/617 (99.8%) patients using the LMA Supreme. Insertion success ranged from 86-100% of patients on first attempt. Insertion times were reported in six studies but were poorly defined in four studies.11,12,24,28 In studies where the end points of insertion times were defined, they were defined either as time to connection of breathing system25 or as time to first ventilation.27 The oropharyngeal leak pressure observed in four studies ranged from 23-39 cm H2O.12,24,25,27 There was a low incidence of complications, including sore throat, vomiting, pain on swallowing, and upper airway trauma. Blood staining was recorded in two studies and was infrequent, 10/140 patients (7%).25,28

Use of the LMA Supreme in difficult airway/emergent situations

Four case reports have shown successful insertion using the LMA Supreme in difficult airway situations. One case involved insertion of the LMA Supreme in a patient undergoing elective surgery,29 and another case involved a patient with ventricular fibrillation requiring chest compression.30 The remaining two cases documented successful insertions in patients whose lungs were difficult to ventilate due to partially obstructed airways.31,32 One LMA Supreme insertion was later replaced with a tracheostomy tube due to an anticipated difficult endotracheal intubation.31 The LMA Supreme was also inserted successfully in 57/60 (95%) cases of simulated difficult airway.18 In a simulated difficult airway, use of the LMA Supreme was associated with a faster insertion time compared with the i-gel. Due to a limited number of reports involving the LMA Supreme as a ventilator device in difficult airway situations, there is insufficient data to conclude about its efficacy in such cases; therefore, additional studies are needed.

Use of LMA Supreme as a conduit for tracheal intubation

In a cannot intubate, cannot ventilate or a failed intubation situation, a supraglottic airway, such as the LMA Classic or LMA Supreme, can be used as a temporary rescue airway and can also serve as a conduit for intubation.33,34 Due to the limited airway luminal diameter of the LMA Supreme, it is difficult to pass an endotracheal tube of adequate size directly through it. Thus, a variety of techniques, mainly via the use of introducers or catheters, have been developed to achieve tracheal intubation with the LMA Supreme. There are four reported techniques using an Aintree Intubation Catheter,35,36 a guidewire-catheter,36,37 a gum elastic bougie,38 or a small endotracheal tube.39

The first technique, which utilizes the Aintree Intubation Catheter, was described by Joffe et al. in a mannequin setting with 20 participants.36 The Aintree Intubation Catheter (4.7-mm internal diameter, 6.3-mm outer diameter) was mounted over a flexible bronchoscope (Fig. 2). Once mounted, the bronchoscope-Aintree assembly was then inserted through the LMA Supreme into the trachea where mean time to completion was 54 (21) sec. After the bronchoscope and LMA Supreme were withdrawn, an endotracheal tube was railroaded over the Aintree Intubation Catheter. There were two failures; in one case, the Aintree Intubation Catheter was dislodged during removal of the LMA Supreme, and in the other case, the bronchoscope-Aintree assembly failed to maneuver past the vocal cords.

Fig. 2
figure 2

(Top) Aintree Intubation Catheter is placed over a flexible bronchoscope. (Bottom) An Aintree Intubation Catheter/bronchoscope complex is inserted through a LMA Supreme

The second technique, utilizing a guidewire, was described in three patients in a mannequin setting.36,37 A guidewire was inserted through the lumen of the bronchoscope, and then both were inserted through the LMA Supreme into the trachea. After removal of the bronchoscope and LMA Supreme, the guidewire was exchanged for an Arndt Exchange Catheter. Subsequently, an endotracheal tube was railroaded over the Arndt Exchange Catheter. In the report by Joffe et al., 36 they compared tracheal intubation with the Aintree Intubation Catheter vs intubation with the Arndt Exchange Catheter Set. Twenty experienced anesthesiologists used an LMA Supreme to perform each method once on an AirSim airway mannequin. Intubation was 90% successful using an Aintree Intubation Catheter and 100% successful using the guidewire. However, faster intubation was achieved with the Aintree Intubation Catheter technique vs the guidewire [54 (21) sec vs 98 (23) sec, respectively; P < 0.0001].

In the third technique, which involved four patients with a normal airway who were undergoing elective surgery, Mathes et al. 38 described insertion of a gum elastic bougie in combination with a flexible bronchoscope through the respiratory lumen of the LMA Supreme. The bronchoscope and LMA Supreme were then removed, and an endotracheal tube was railroaded over the bougie into the trachea. Success was achieved in two of the four patients. In a variation of this technique, a nasogastric tube was used as an adjunct for tracheal intubation through the LMA Supreme in four patients with a normal airway who were undergoing elective surgery. The bronchoscope was first inserted into a nasogastric tube before passing through the LMA Supreme. The bronchoscope was then exchanged for a bougie, followed by removal of the LMA Supreme. An endotracheal tube was then railroaded over the nasogastric tube/bougie into the trachea. Success was achieved in all four cases.

Lastly, in the fourth technique described by Carron et al.,39 a small endotracheal tube was used as an intermediary for a larger tube in a patient with three failed attempts at laryngoscopic tracheal intubation. A size 4 LMA Supreme was first inserted for ventilation. A small endotracheal tube (6.0-mm internal diameter) was then mounted over a bronchoscope and inserted through the LMA Supreme into the trachea. The bronchoscope and LMA Supreme were withdrawn, and a 4.0-mm Cook Airway Exchange Catheter was inserted into the small endotracheal tube. Subsequently, the small endotracheal tube was removed and a 7.5-mm internal diameter endotracheal tube was railroaded into the trachea.

The above reports suggest that the LMA Supreme in combination with introducers/catheters can be a viable alternative to achieve tracheal intubation. However, there are limited data to support the efficacy of this device as a conduit for tracheal intubation in difficult intubation situations.

Limitations

There are a number of limitations which affect the interpretation of these data. It is difficult to generate solid conclusions with a limited number of RCTs with sufficient power to compare the LMA Supreme with other airway devices. Heterogeneity in the definition of outcomes, including airway device insertion time, makes it difficult to draw comparisons between studies. In addition, there are few studies documenting the use of the LMA Supreme in difficult airway situations or as a conduit for intubation. Thus, additional RCTs are needed.

Conclusions

The LMA Supreme is a single-use version of the reusable double-lumen LMA Proseal. With the built-in pre-shaped shaft, it can be inserted in the same fashion as the LMA Fastrach. The efficacy and safety of the LMA Supreme as a stand-alone supraglottic airway has been shown in a number of studies. Insertion success, insertion time, and incidence of complications were shown to be comparable with the LMA Proseal in patients with normal airways. In several studies, the LMA Proseal was shown to have higher oropharyngeal leak pressure than the LMA Supreme. Furthermore, the LMA Supreme has been used successfully in a limited number of difficult airway cases. The LMA Supreme can be used as a conduit for tracheal intubation through the use of introducers or catheters. Definitive outcome-based clinical trials that are conducted meticulously, powered appropriately, and reported transparently are required to compare the efficacy of the LMA Supreme with other supraglottic airway devices in both elective and emergency airway management settings.