Abstract
We describe a new catheter for emergency ventilation of patients difficult to intubate. This catheter can be inserted through the crico-thyroid membrane or the first or second intertracheal ringspace with ease in an almost atraumatic fashion. The outside part of the device, with its dual attachment system, can be connected to conventional resuscitation equipment by its 15 mm male end or to a high-pressure oxygen source by its luer-lock fitting. A double angulation maintains the intratracheal portion of the catheter in the axis of the trachea and the external part in close contact with the larynx. A Velcro band attached to two lateral flanges keeps the catheter in place. The results of transtracheal catheterizalion of 48 patients by means of this new device are discussed.
Résumé
Un nouveau système de canulation transtrachéale ou transcrico-thyroidienne est présenté. Les multiples avantages qu’il procure sont a) sa mise en place rapide, b) sa connexion aisée, par son embout ISO 15 mm mâle etlou par son embout luer-lock femelle scellé en son milieu, à tout circuit d’anesthésie ou de réanimation, c) ses 2 ailettes, permettant une fixation solide autour du cou du patient et d) sa double angulation, assurant une position parallèle à la trachée et un positionnement adéquat contre le larynx du patient.
Un total de 4S patients ayant été ventilés avec succès au moyen de ce cathéter est présenté. Les indications étaient les suivantes: a) patients difficiles à intuber: huit cas b) patients ayant déjà une fixation intermaxillaire associée à une fracture de la base etlou de la lame criblée; dix cas c) microlaryngoscopies en suspension avec laser: vingt cas d) endoscopies: dix cas.
Vu les excellents résultats obtenus et la facilité d’administrer de l’oxygène par cette voie, on ne peut que souligner l’ importance d’avoir rapidement sous la main un cathéter transtrachéal et des connexions adéquates.
Article PDF
Similar content being viewed by others
References
Jacoby JJ, Hamelberg W, Reed JP, Gillespie B, Hitchcock FA. Simple method of artificial respiration. Am J Physiol 1951; 167: 798–9.
Jacoby JJ, Hamelberg W, Ziegler CH, Flory FA, Jones JR. Transtracheal resuscitation. JAMA 1956; 162: 625–8.
Sanders RD. Two ventilating attachments for bronchoscopes. Del Med J 1967; 39: 170–5.
Spoerel WE, Narayanan PS, Singh NP. Transtracheal ventilation. Br J Anaesth 1971; 43: 932–9.
Jacobs HB. Emergency percutaneous transtracheal catheter and ventilator. J Trauma 1972; 12: 50–55.
Jacobs HB. Needle-catheter brings oxygen to the trachea. JAMA 1972; 222: 1231–3.
Nunn JF. Applied Respiratory Physiology. London: Butterworth & Co., 1969: 146–7.
Garden E, Becker G, Hamood H. Percutaneous jet ventilation. Ann Otol Rhinol Laryngol 1976; 85: 652–5.
Smith BR, Babinski M, Klain M, Pfaeffie H. Percutaneous transtracheal ventilation. JACEP 1976; 5: 765–70.
Stinson TW. A simple connector for transtracheal ventilation. Anesthesiology 1977; 47: 232.
DunlapLB. A modified, simple device for the emergency administration of percutaneous transtracheal ventilation. JACEP 1978; 7: 42–6.
Fisher JA. A “last ditch” airway. Can Anaesth Soc J 1979; 26: 225–30.
Carlton DM, Zide MF. An easily constructed cricothyroidotomy device for emergency airway management. J Oral Surg 1980; 38: 623–4.
Delisser EA, Muravchick S. Emergency transtracheal ventilation. Anesthesiology 1981; 55: 606–7.
Hilton PJ. A simple connector for cricothyroid cannulation. Anaesthesia 1982; 37: 220.
Scuderi PE, McLeskey CH, Comer PB. Emergency percutaneous transtracheal ventilation during anaesthesia using readily available equipment. Anesth Analg 1982; 61: 867–70.
Gildar JS. A simple system for transtracheal ventilation. Anesthesiology 1983; 58: 106.
Aye LS. Percutaneous transtracheal ventilation. Anesth Analg 1983; 62: 619.
Pate lR. Systems for transtracheal ventilation. Anesthesiology 1983; 59: 165.
Ponecher T, Bing J, Cuby C, Fleck G, Coppe B, Otteni JC. Ventilation translaryngée de sauvetage par aiguille de Tuohy; utilisation en cas d’impossibilité d’intuber et de ventiler un patient curarisé. Ann Fr Anesth Réanim 1984; 3: 54–8.
Klain M, Smith BR. High frequency percutaneous transtracheal jet ventilation. Crit Care Med 1977; 5: 280–7.
Klain M, Miller J, Kalla R. Emergency use of high frequency jet ventilation. Crit Care Med 1981; 9: 160.
Klain M. High frequency ventilators and methods of their use. Presented at Critical Care Medicine 20th Annual Symposium 1982, Las Vegas (Nevada).
Basset JM, Eurin B, Francois M, Hertzog C, Laquerriere MC, Ardoin C La ventilation à haute fréquence par voie intercricothyroïdienne dans les endoscopies ORL. Notre expérience de 83 cas. Ann Oto Laryng 1982: 99: 159–66.
Patel KF, Hicks JN. Prevention of fire hazards associated with use of carbon dioxide lasers. Anesth Analg 1981; 60: 885–8.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Ravussin, P., Freeman, J. A new transtracheal catheter for ventilation and resuscitation. Can. Anaesth. Soc. J. 32, 60–64 (1985). https://doi.org/10.1007/BF03008540
Issue Date:
DOI: https://doi.org/10.1007/BF03008540